MEDIVATORS REPROCESSING SYSTEMS
9
3 SETUP & FAMILIARIZATION
Device Installation
SCOPE BUDDY can only be used with the power supply supplied with the device. Do not attempt to
substitute or use a non-approved replacement power supply.
The cycle run time can be adjusted by the operator, but always verify the rinse time is appropriate for
the endoscope being flushed.
Do not use alcohol or alcohol-based products, bleach solutions, or hydrogen peroxide in the SCOPE
BUDDY.
The tubing hookups, hookup extensions, and fluid intake line are not autoclavable and must be
reprocessed using low temperature decontamination methods only.
SCOPE BUDDY is designed to be placed on a counter top or mounted to the wall by using the
supplied wall-mounting bracket. If mounting to a wall, ensure the bracket is properly secured using
the correct mounting hardware, adequate to support at least ten pounds of weight.
SCOPE BUDDY must be located within a vertical height no greater than three feet above the
incoming detergent or rinse water solution. The distance between the top of the fluid solution and
SCOPE BUDDY must not exceed three feet.
SCOPE BUDDY is designed to operate in a damp environment, but should not be immersed in any
fluids. Do not use the device if it becomes immersed in any type of fluid, or if fluid is observed leaking
from the device.