SCOPE BUDDY USER/SERVICE MANUAL
14
Programming Time to Flush 100 mls through Pentax Endoscopes including those with an Elevator Wire
Channel (serial number 201345)
Note:
When flushing the Forward Water Jet of a Pentax endoscope, always ensure the forward
water jet adapter does not contain a check valve. Only a forward water jet non-check valve
adapter can be used. Also, the Forward Water Jet can be flushed simultaneously with the other
endoscope channels, and does not require the use of the Elevator Wire Channel Hook-up. The
Pentax forward water jet can be flushed using the DSD Hook-up or the All-channel Irrigator
Hook-up.
Small Caliber Endoscopes 35 seconds
Gastroscopes (including forward water
jet w/non-check valve adapter)
35 seconds
Duodenoscopes (including elevator
wire channel)
1 minute 5 seconds
Colonoscopes (including forward water
jet w/non-check valve adapter)
1 minutes 5 seconds
Sigmoidoscopes 35 seconds (using the forward
water jet w/non-check valve
adapter
Ultrasound Scope 35 seconds
Programming Time to Flush 100 mls through Fujinon Endoscopes excluding those with an Elevator Wire
Channel or a Forward Water Jet (serial number 201345)
Small Caliber Endoscopes 35 seconds
Gastroscopes 1 minute 30 seconds
Gastroscope: Model EG-450HR 35 seconds
Duodenoscopes (including elevator
wire channel)
1 minute 5 seconds
Duodenoscopes (elevator wire channel
only, 15 mls)
3 minutes (using EWC hook-up)
Colonoscopes 1 minutes 5 seconds
Sigmoidoscopes 1 minute 5 seconds
Ultrasound Scope 1 minute 5 seconds
Fjuinon Forward Water Jet 1 minute 30 seconds (using EWC
hook-up)
Elevator Wire Channel – Fujinon Endoscopes
SCOPE BUDDY can be used to flush an endoscope’s elevator wire channel (EWC). Due to the internal
structure of the Fujinon EWC channel, it must be flushed independently of the other endoscope channels
The Fujinon endoscope’s EWC simultaneously with other endoscope channels as the elevator wire channel will
not be properly flushed per the endoscope manufacturer’s standards and guidelines. The independent flushing
of the elevator wire channel using Scope Buddy’s Elevator Channel Hook-up ensures fluid is delivered directly
to this channel. Go to the section labeled “Endoscope Flushing using the Elevator Wire Channel Hook-up –
Fujinon Endoscopes” for further instructions.
.