SCOPE BUDDY USER/SERVICE MANUAL
24
Purging an Endoscope of Detergent:
1. Remove the Fluid Intake Line from the detergent solution and place on the counter.
2. Press the START/STOP button once to reset the timer to the original pre-set time amount.
3. Press the START/STOP button again to begin pumping. The timer display will now begin counting
down as some of the detergent solution is pumped out of the endoscope. Verify fluid is exiting the distal
tip of the endoscope.
4. Press START/STOP button to halt pumping when fluid no longer exits the endoscope. Note: Most fluid
is now purged from the endoscope, but some residual fluid will remain in the endoscope.
Rinsing an Endoscope with Water:
1. Dispose contaminated detergent solution and thoroughly rinse the sink.
2. Fill the sink or the fluid container with clean rinse water and place the Fluid Intake Line into the
clean water.
3. Press the START/STOP button once to reset the timer to the original pre-set time amount.
4. Press the START/STOP button again to begin the cycle; the timer display will now begin counting
down. Rinse water will now be drawn up, pumped through SCOPE BUDDY, and will flow out of the
endoscope channels .
5. Verify rinse water is exiting all endoscope channels. If fluid fails to come out of the endoscope, then the
endoscope may need to be manually brushed and cleaned again. The flushing process will then need
to be repeated until flow through the endoscope is verified.