MEDIVATORS REPROCESSING SYSTEMS
29
Caution: The screened end of the Fluid Intake Line must remain deeply submerged in
solution to ensure proper fluid delivery and no interruption in fluid flow.
Note: A flushing cycle can be paused by pressing the START/STOP button. Pressing the
START/STOP button again will resume the original cycle. If, however, an arrow key was
pressed during the pause cycle, then pressing the START/STOP button will begin a cycle
from the original pre-set time amount.
6. Once the timer counts down to zero, fluid flow will stop and a five-second audible tone will be heard.
During this time, the timer display will flash all zeros “00.00” and then remain with a zero display
providing an indication that the cycle is complete.
Purging an Endoscope of Rinse Water
1. Remove the Fluid Intake Line from the rinse water and place on the counter.
2. Press the START/STOP button once to reset the timer to the original pre-set time amount.
3. Press the START/STOP button again to begin pumping. The timer display will now begin counting
down as some of the rinse solution is pumped out the endoscope. Verify fluid is exiting the endoscope
channel.
4. Press START/STOP button to halt pumping when fluid no longer exits the endoscope.
Note: Residual fluid will remain in the endoscope. Do not use SCOPE BUDDY to deliver
Alcohol to the endoscope. Alcohol should never be used in SCOPE BUDDY.
5. Upon successful completion of these steps, the endoscope is now ready to be placed in an Automated
Endoscope Reprocessor, or it can be manually reprocessed for high-level disinfection.