PERFECT page 19
10.1 Closure commands for ramp segments
NEXT Follow-on with next programme segment
SPWT (T)
SET-POINT WAIT
Wait until the setpoint temperature is
reached.
The oven only starts the next programme
segment when the programmed setpoint
temperature has been reached, even if the
programmed heating time has already
elapsed.
LOOP Ramp repeat function
The set programme is repeated after passing
through all programmed segments.
1-99 = repeats
cont = continuous repeat function
HOLD End of programme without switching off the
heating; temperature and all other settings
(e.g. air valve) are maintained.
END End of programme, heating is switched off,
all other settings (e.g. air valve) are reset to
base status.
The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.
°C
Ramp
closure
command
ramp
segment 1
spwt (t)
t=time
segment 1
closure
command
ramp
segment 2
next
closure
command
ramp
segment 4
next
closure
command
ramp
segment 5
end
closure
command
ramp
segment 3
spwt (t)
segment 3
segment 1
segment 5
segment 4
delayed
programme start