Defrosting (ICP incubators only)
Setting: off 3h 6h 12h 24h 48h
(see Section: “Automatic defrosting on ICP incubators”)
Factory setting: 12h
Tolerance margin ASF
Range:
Universal ovens Uxx 2 to 20
Sterilisers Sxx 2 to 20
Incubator Ixx 0.5 to 5
(see Section: „Temperature monitor”)
Temperature monitor function
Adjustable temperature monitor (TWW)
Protection Class 3.1 to DIN 12 880
Adjustable temperature limiter (TWB)
Protection Class 2 to DIN 12 880
(see Section: „Temperature monitor”)
Language
Settings: GERMAN, English, franCEAIS, ESPANOL and italIANO
Calibration correction for user-calibration
CAL1 to CAL3
Adjust - oven-temperature for adjustment
RE-Adjusting temperature value
(see Section: „Calibration“)
Exit setup mode
= store all settings and exit setup mode.
12.1 Real-time clock
The real-time clock is set in SETUP and includes date and clock time.
The real-time clock serves for documentation according to GLP.
Date and clock time are marked in the report print.
On the graphics print the time axis is marked in real-time.
The clock runs with a buffer battery independently of the mains power supply.
The built-in lithium battery Type CR 2032 has a life of approx. 10 years.