21.1 Cleaning IPP incubators
In order to ensure the proper function and long life of the Peltier cooling module it is essential to remove
any dust deposits from the heat sink on the back of the incubator (using vacuum cleaner, brush or bottle
brush depending on the amount of dust).
To assist with cleaning, the protective cover can be removed after releasing the screws.
21.2 Cleaning ICP incubators
In order to ensure the proper function and long life of the refrigeration unit it is essential to remove any
deposits from the condenser 1 (using vacuum cleaner, brush or bottle brush depending on the amount of
dust).
WARNING! Before opening the housing, always pull out the supply plug!
Release the screws 2; the front ventilation grill 3 can then be removed so that the condenser can be
cleaned.
After releasing the screws 4 the rear ventilation grill can also be removed so that the refrigeration unit can
be cleaned from both sides if there is a large amount of dirt.