PERFECT page 8
Do not place the oven on a readily inflammable support surface!
WARNING! Always pull out the supply plug before
opening the oven cover!
3.3 Electrical equipment
Operating voltage see label 50/60 Hz
Current rating see label
Protection Class 1, i.e. operating isolation with ground connection to EN 61 010
Protection IP20 to EN 60 529
Interference suppression to EN55011 Class B
Oven protected by a fuse 250V/15A fast blow
Controller protected by a 100 mA fuse (200 mA on 115 V)
When connecting a MEMMERT oven to the electrical supply you have to observe any local
regulations which apply (e.g. in Germany DIN VDE 0100 with FI protection circuit)
This product is intended to operate on a supply network with a system impedance Zmax at the
transfer point (building connection) of 0.292 Ohm max. The user has to ensure that the product is
only operated on an electrical supply network which meets these requirements. If necessary, details
of the system impedance can be obtained from the local electricity supply authority.
3.4 External connection
Equipment connected to the external connections must have interfaces which meet the requirements for
safe low voltage (e.g. PC, printer).