Note:
If all calibration corrections are set to 0.0°C the factory calibration is restored.
Calibration correction
Range –9.9°C to +9.9°C
Calibration temperature
Range down to 10°C below
CAL2
Calibration point 1
Calibration correction
Range –9.9°C to +9.9°C
Calibration temperature
Range 10°C above CAL1 to
10°C below CAL3
Calibration point 2
Calibration correction
Range –9.9°C to +9.9°C
Calibration temperature Range
10°C above CAL2 up to nominal
temperature
Calibration point 3