CL-Profiler operating instructions page 14 of 15
www.cl-profiler.com
After connecting the second probe on your PC you will see two needles
(red and green) in your measure window.
You can adjust the radial position of each probe separately, but it is not
necessary that both probes are exact identical. You can simply activate
“zero startpoint” before you start the profiling. The software will then
place both probes on the same level independent of the physical position
of the Ruby tip positions.
Please note for wide grooves or/and mica slots that you use the 8mm ruby
tips on both probes.
V) Maintenance
Make sure the cable to the sensor and USB port is not damaged. Clean
sensor when necessary.
It is recommended that the sensor is calibrated every two years.
VI) Warranty
The warranty covers construction and material faults of the CL-profiler
equipment.
The warranty period is valuable for 1 year starting on the invoice date.
Damage due to incorrect or inexpert handling is excluded from the war-
ranty.
The CL-profiler probe must be placed as radial as possible especially on
commutators. Angled probe position, can cause to damage.
The ballpoint tip diameter must be:
- for mica slots: approximately 4 times larger then the full width of
the mica slot including the chamfers.
- for spiral grooves: approximately 3 times larger then the full width
of the groove including the chamfers.