EasyManuals Logo

MessKonzept FTC320 User Manual

MessKonzept FTC320
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
Page 42 of 47
Chapter 9
Appendix: Specifications
9.1 Specification of Thermal Conductivity Measurement
Attribute Range / Precision
Linearity < 1 % of range
Warm up time
Approx. 20 min; up to 1 h for
small measuring ranges
Flow rate
10 l/h - 150 l/h,
60 l/h - 80 l/h (recommended)
T90-time
< 1 sec at flow rate higher 60 l/h
(or dependent on user selected
T-90-filter time)
Noise < 0.5 % of smallest range
Drift at zero point
< 2 % of smallest range per
week
Repeatability < 1 % of range
Error due to change of ambient temper-
ature
< 1 % of smallest range per
10 K temperature change
Error due to change of flow at 80 l/h
< 1 % of smallest range per
10 l/h
Error due to change of pressure
(800 hPa < p < 1200 hPa)
< 1 % of smallest range per
10 hPa
Table 9.1: Specification of TC measurement
File name:
FTC320Ex Operating Manual_1.09KD201009MPO3V04.pdf

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MessKonzept FTC320 and is the answer not in the manual?

MessKonzept FTC320 Specifications

General IconGeneral
BrandMessKonzept
ModelFTC320
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals