MI 2087 AL2 EasiTEST Measurement instructions
32
Notes:
q Nominal input voltage range is 100 V ÷ 440 V. If input voltage is out of the
range, the voltage value equipped with “!” mark will be displayed after pressing
START key. Warning sound is also emitted.
q If test result is out of display range, “>1999 Ω” message will be displayed!
q Specified accuracy of tested parameters is valid only if mains voltage is stabile
during the measurement!
q “bat” message displayed during or after finishing the measurement means
batteries are too weak to guarantee correct result. Replace the batteries