MI 2087 AL2 EasiTEST Other operations
33
4 Other operations
4.1 Reset of the instrument
If any malfunction is noticed when dealing with the EasiTest, it is advisable to RESET
the instrument. In that case all settable parameters will be set to their initial values;
see the table below. Stored results will not be erased.
How to reset the instrument
q Switch OFF the instrument.
q Press the CAL key and keep it pressed, while switching ON the instrument.
rES” message will be displayed for a while meaning RESET function has been
completed.
Parameter Function Initial value
Compensation of test leads Continuity Annulled
Nominal differential current All RCD functions 30 mA
Type of RCD All RCD functions G (general)
Limit contact voltage All RCD functions 50 V
Multiplier of nominal
differential current
tN ×1/2
Test voltage Insulation 250 V
Table 2: Initial values of settable parameters