MI 3109 EurotestPV Lite Technical specifications
66
9.2 Continuity
9.2.1 Resistance R LOW
Measuring range according to EN61557 is 0.16 1999 .
(3 % of reading + 3 digits)
Open-circuit voltage ........................... 6.5 VDC 9 VDC
Measuring current .............................. min. 200 mA into load resistance of 2
Test lead compensation ..................... up to 5
The number of possible tests ............ > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
9.2.2 Resistance CONTINUITY
(5 % of reading + 3 digits)
Open-circuit voltage ........................... 6.5 VDC 9 VDC
Short-circuit current ........................... max. 8.5 mA
Test lead compensation ..................... up to 5