EasyManua.ls Logo

METREL MD 9273 - Technical Specifications; General specifications - Environment conditions; Measurement specifications

METREL MD 9273
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MD 9273 Leakage Clamp Technical Specifications
23
4 Technical Specifications
4.1 General specifications Environment conditions
Measurement category clamp-meter
CAT III 600V
Measurement category test lead
CAT III 600V
Maximum current
2
100 A
Pollution degree
2
Operation
Indoor
Altitude
< 2000 m
Operation temperature
0 ºC 40 ºC (<80 % H.R., <10 ºC without
condensation)
Storage temperature
10 ºC – 60 ºC (<70 % H.R., extract batteries)
Temperature coefficient:
0.1 (of specified precision) / ºC (<18 ºC or >28 ºC)
Functioning principle:
Sampling the signal at high speed
Data update on screen:
Twice a second
Screen:
3 3/4 + 3 digits LCD screen
Range selection:
Automatic and manual selection
Over-range indication:
-0L- is shown on the LCD screen, fast beep
Polarity indication:
-“ is shown automatically
E.M.C.:
Emmission: Class B
Immunity: Portable test and measurement
equipment
Meets EN 61326-1:2013, EN 61326-2-2:2013
Safety:
Meets EN 61010-1:2010, EN 61010-2-32:2012,
EN 61010-2-033:2012, EN 61010-031:2015
Bluetooth®:
v3.0
Power supply:
3 V
Type of batteries:
2 x 1,5 V AA alkaline batteries
Consumption:
5 mA average, depending on function.
50 mA average in Bluetooth® mode
Duration of the batteries:
200 - 400 h (batteries 2000 mAh)
20 40 h (Bluetooth® mode)
Dimensions:
185 mm x 62 mm x 42 mm
Weight:
Approx. 210 g (without battery)
4.2 Measurement specifications
Current and voltage support automatic ranging within the same function.
2
This equipment cannot be exposed to currents above the limit

Table of Contents

Related product manuals