Calibrateur multifonctions CX 1651
172 Manuel de l'utilisateur
1. Calibration of industrial thermometers with Pt100 sensor and 20 mA / 10 V output:
Function setup: Calibrator T RTD function
OUTPUT 140-41 ON
Multimeter DCV or DCI depending on the output signal type of the unit
being tested
Adapter Opt. 140-41
2. Calibration of industrial thermometers with thermocouple sensor and frequency output:
Function setup: Calibrator T TC function
OUTPUT 140-41 ON
Multimeter F
Adapter Opt. 140-41
3. Calibration of a single phase “electrical power/current loop” transducer
Function setup: Calibrator P function (voltage, current, power factor, frequency)
Multimeter DCI
Adapter Opt. 40
DUT
INPUT OUTPUT
Pt100 H L
DUT
INPUT OUTPUT
TC H L
INPUT OUTPUT
HU LU HI LI +I -I
DUT
power transducer
Opt. 40 adapter