For pre-rinsing, you will need the holder for the eluent aspiration fil-
ter (6.2744.360), a syringe and the purge needle (6.2816.040).
This procedure can also be found in the video "Inserting a new aspi-
ration filter" at the following link: ic-help.metrohm.com
2
Mounting aspiration filter
NOTE
Always wear gloves when handling the aspiration filter.
■ Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.
■ Tighten the aspiration filter to the filter holder.
Figure 4
Installing tubing weighting and aspiration filter