8.5 Safety specifications
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
Wavelength preci-
sion
Based on one
instrument
< 0.005 nm
Based on an
instrument
group
< 0.02 nm
Photometric noise
400 to 700 nm < 50 mAU
700 to 2,500
nm
< 20 mAU
8.5 Safety specifications
This instrument fulfills the following electrical safety requirements:
CE marking in accordance with the EU directives:
2006/95/EC (Low Voltage Directive, LVD)
2004/108/EC (EMC Directive, EMC)
Design and testing According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1, protection
class I, EN/IEC 61010-2-010, EN/IEC 60529, degree of protection IP65.
Safety instructions This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.
8.6 Standards and directives
The instrument complies with the following directives:
Directive 94/62/EC (packaging and packaging waste)
Directive 2002/95/EC (RoHS 1)
Directive 2002/96/EC (waste electrical and electronic equipment)
Regulation 1907/2006/EC (REACH)