3 Installation
■■■■■■■■
15
3.7 Setting up accessories
Metrohm supplies various sample cups for your NIRS DS2500 Analyzer:
Samples Sample cups Order number
Solid samples, powders NIRS mini sample cups, 10 pcs., incl. 100 dis-
posable backs
6.7402.030
Solid samples, granulated substances NIRS DS2500 large sample cup 6.7402.050
Liquid samples and suspensions NIRS liquid sample kit transflection 6.7400.010
3.7.1 Inserting the sample cup for granulated substances
Use the NIRS large sample cup (6.7402.050) for measuring coarse solid
samples (granulated substances).
Inserting the sample cup for granulated substances
Required accessories
NIRS DS2500 large sample cup (6.7402.050)
1
Clean the sample cup and the sampling window with a lens cleaner
cloth.
2
Fill the granulated substance into the sample cup.
Make sure that the glass bottom of the sample cup is completely
covered and that the layer of sample is at least 1 cm thick.
3
Place the sample cup centrally onto the sampling window.

Questions and Answers: