EasyManua.ls Logo

Metrohm NIRS DS2500 - Page 22

Metrohm NIRS DS2500
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.6 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
3
In the software, start Reference Standardization. Follow the
instructions in the software.
When the software prompt appears, insert the reflection standard
(R8012007) into the opening of the holder. The label of the
reflection standard must be visible.
Close the lid.
Wait until Reference Standardization is completed.
Remove the reflection standard from the holder.
Regulated industries
If you put the NIRS DS2500 Analyzer into operation in a regulated indus-
try, then additional tests are required after the procedures described in the
chapter "Non-regulated industries" have been performed.
In regulated industries, the Vision Pharma software version and the NIRS
reflection standard, 7 pcs. (6.7450.010) are required for the instrument to
be put into operation.
Information on which additional tests are required for your specific regu-
lated industry can be found in the applicable standards.

Related product manuals