Cermet II User’s Manual
Michell Instruments 25
MAINTENANCE
5 MAINTENANCE
Calibration
Routine maintenance of the Cermet II Transmitter is con ned to regular re-calibration
by exposure of the transmitter to sample gases of known moisture content to ensure
that the stated accuracy of the transmitter is maintained. Calibration services traceable
to the UK
National Physical Laboratory
(NPL) and the US
National Institute of Standards
and Technology
(NIST) are provided by Michell Instruments.
Michell Instruments offers a variety of re-calibration and exchange transmitter schemes
to suit speci c needs. A Michell representative can provide detailed, custom advice (for
Michell Instruments’ contact information go to www.michell.com).
Transmitter Guard Replacement
The transmitter is supplied with a white HDPE guard (standard) or a stainless steel
guard (if speci ed at time or order). The method of replacement is the same for both
types.
HDPE Guard
The HDPE guard provides <10μm protection to the dew-point sensor. It is designed
to show any contamination and the guard should be changed if the surface becomes
discolored.
When replacing the guard, care should be taken to handle the guard by the bottom part
only. Replacement guards (EA2-HDPE) - pack of 10 - can be obtained by contacting
Michell Instruments (www.michell.com) or your local distributor.
t
e
r
0
9
0
6
M
I
C
H
E
I
n
s
t
r
u
m
e
n
x
HANDLE,
USING
GLOVES, BY
BLACK PART
ONLY
Figure 21
Replacement of HDPE Guard
Stainless Steel Guard
The stainless steel guard provides <80μm protection to the dew-point sensor. It is
designed to show any contamination and the guard should be changed if the surface
becomes discolored.
When replacing the guard, care should be taken to handle the guard by the bottom part
only. A replacement guard (SSG) can be obtained by contacting Michell Instruments
(www.michell.com) or your local distributor.
Bonded Seal
If the installed bonded seal gets damaged or lost, a pack of 5 replacement bonded seals
can be obtained by contacting Michell Instruments, or your local distributor, and quoting
part number BS-58-PK5.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Michell Instruments Cermet II and is the answer not in the manual?

Michell Instruments Cermet II Specifications

General IconGeneral
BrandMichell Instruments
ModelCermet II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals