EasyManua.ls Logo

MICRO-EPSILON thermoMETER CT - Page 87

MICRO-EPSILON thermoMETER CT
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 87
Emissivity
thermoMETER CT
14.3 Characteristic Emissivity
In the case that none of the methods mentioned above help to determine the emissivity you may use the
emissivity tables, see Chap. A 2, see Chap. A 3. These are only average values. The actual emissivity of a ma-
terial depends on the following factors:
- Temperature
- Measuring angle
- Geometry of the surface (smooth, convex, concave)
- Thickness of the material
- Constitution of the surface (polished, oxidized, rough, sandblast)
- Spectral range of the measurement
- Transmissivity (e.g. with thin films)