Problem Possible causes Recommended actions
Inaccurate referred
density reading
Inaccurate density measurement
Inaccurate temperature measurement
Inaccurate pressure value
Incorrect reference conditions
Incorrect API table selection
Verify the line density value. If it is not ac-
curate, see Section 10.7.
Verify the line temperature value. If it is
not accurate, see Section 10.8.
Ensure that the application is configured to
use the appropriate temperature source.
Ensure that the pressure source is config-
ured correctly, that the external pressure
device is operating correctly, and that both
devices are using the same measurement
units.
Ensure that reference temperature and ref-
erence pressure, if applicable, are config-
ured correctly.
Ensure that the selected API table is appro-
priate for the process fluid.
10.10 Concentration measurement problems
Problem Possible causes Recommended actions
Significantly incorrect
concentration meas-
urement after loading
matrix
The wrong temperature or density unit
was configured when the matrix was loa-
ded
Set the temperature and density units to
the units used when the matrix was built,
then reload the matrix. For custom matri-
ces, contact Micro Motion.
Inaccurate concentra-
tion measurement
reading
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect matrix data
Inappropriate trim values
Verify the line density value. If it is not ac-
curate, see Section 10.7.
Verify the line temperature value. If it is
not accurate, see Section 10.8.
Ensure that the application is configured to
use the appropriate temperature source.
Ensure that reference temperature is con-
figured correctly.
Ensure that the appropriate matrix is ac-
tive.
Ensure that the matrix is configured cor-
rectly.
Ensure that the matrix curve fit is good.
Adjust the extrapolation limits for the ac-
tive matrix.
Adjust measurement with a concentration
offset trim. See Section 9.5.
Troubleshooting
152 Micro Motion
®
Compact Density Meters (CDM)