10.11 Velocity measurement problems
Velocity measurement problems and recommended actionsTable 10-4:
Problem Possible causes Recommended actions
Non-zero velocity
reading at no-flow
conditions or at zero
offset
Misaligned piping (especially in new instal-
lations)
Open or leaking valve
Incorrect sensor zero
Zero the meter. See Section 2.4.
Check for open or leaking valves or seals.
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
Contact Micro Motion.
Erratic non-zero ve-
locity reading at no-
flow conditions
Leaking valve or seal
Two-phase flow
Plugged or coated sensor tube
Incorrect sensor orientation
Wiring problem
Vibration in pipeline at rate close to sensor
tube frequency
Damping value too low
Mounting stress on sensor
Verify that the sensor orientation is appro-
priate for your application (refer to the
sensor installation manual).
Check the drive gain and the pickoff volt-
age. See Section 10.24 and Section 10.25.
Purge the sensor tubes.
Check for open or leaking valves or seals.
Check for sources of vibration.
Verify damping configuration.
Verify that the measurement units are con-
figured correctly for your application.
Check for two-phase flow. See
Section 10.23.
Check for radio frequency interference.
See Section 10.20.
Contact Micro Motion.