Default settings for mA Output Process VariableTable 6-1:
Device Channel mA output
Default process variable assign-
ment
CDM mA Channel A Primary mA output Density
Channel B Secondary mA output Temperature
CDM DO Channel A Primary mA output Density
CDM TPS Channel A Primary mA output Temperature
Postrequisites
If you changed the setting of mA Output Process Variable, verify the settings of Lower Range
Value (LRV) and Upper Range Value (URV).
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several
application-specific options. Different communications tools may use different labels for
the options.
Options for mA Output Process VariableTable 6-2:
Process variable
Label
Display ProLink III Field Communicator
Standard
Line Density DENS Line Density Density
Line Temperature TEMP Line Temperature Temperature
Case Temperature TCASE Case Temperature Case Temperature
Tube-Case Temperature Dif-
ference
TDIFF Tube-Case Temperature Dif-
ference
Tube-Case Temp Difference
Line Temperature (External
or Fixed)
EXT T Line Temperature (External
or Fixed)
External Temperature
Line Pressure (External or
Fixed)
EXTP Line Pressure (External or
Fixed)
External Pressure
Drive Gain DGAIN Drive Gain Drive Gain
Sensor Time Period TP B Sensor Time Period Sensor Time Period
Velocity VEL Velocity Velocity
User-Defined Calculation
Output
UCALC User-Defined Calculation
Output
User-Defined Calculation
Output
API referral
Referred Density (API) RDENS Referred Density (API) Density at Reference (API)
Integrate the meter with the control system
Configuration and Use Manual 85