Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. To save the displayed value to transmitter memory, activate Scroll and Select
simultaneously and hold until the display changes.
If the displayed value is the same as the value in transmitter memory, you will be
returned to the previous screen.
If the displayed value is not the same as the value in transmitter memory, SAVE/
YES? flashes on the display. Activate Select.
6. Switch back from exponential notation to decimal notation.
a. Activate Select until the E is flashing.
b. Activate Select until d is displayed.
c. Activate Select.
B.4 Display codes for process variables
Display codes for process variablesTable B-2:
Code Definition
Standard
DENS Line Density
TEMP Line Temperature
TCASE Case Temperature
TDIFF Tube-Case Temperature Difference
EXT T Line Temperature (External)
EXT P Line Pressure (External)
VEL Velocity
DGAIN Drive Gain
TP B Sensor Time Period
UCALC User-Defined Calculation Output
API referral
RDENS Referred Density (API)
Concentration measurement
SG Specific Gravity
CONC Concentration
Using the transmitter display
176 Micro Motion
®
Compact Density Meters (CDM)

Table of Contents