Transmitter status reported by status LED (continued)Table 10-1:
LED state Description Recommendation
Red One or more high-severity alerts are active. A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
Related information
View and acknowledge status alerts
10.6 Status alerts, causes, and recommendations
Alert num-
ber Alert title Possible cause Recommended actions
A001 EEPROM Error The transmitter has detected a
problem communicating with the
sensor.
Cycle power to the meter.
Contact Micro Motion.
A002 RAM Error The transmitter has detected a
problem communicating with the
sensor.
Cycle power to the meter.
Contact Micro Motion.
A003 No Sensor Response The transmitter is not receiving one
or more basic electrical signals from
the sensor.
Check the drive gain and the pickoff
voltage. See Section 10.24 and
Section 10.25.
Check the integrity of the sensor tubes.
A004 Temperature Over-
range
The RTD resistance is out of range
for the sensor.
Check your process conditions against
the values reported by the device.
Verify temperature characterization or
calibration parameters.
Contact Micro Motion.
A005 Velocity Overrange The measured velocity is greater
than the maximum velocity limit of
the sensor (ΔT greater than 200 µs).
If other alerts are present, resolve those
alert conditions first. If the current alert
persists, continue with the recommen-
ded actions.
Check your process conditions against
the values reported by the device.
Check for two-phase flow. See
Section 10.23.
A006 Characterization Re-
quired
Calibration factors have not been
entered, or the sensor type is incor-
rect, or the calibration factors are
incorrect for the sensor type.
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
Contact Micro Motion.
Troubleshooting
Configuration and Use Manual 143

Table of Contents

Other manuals for Micro Motion CDM100P