Alert num-
ber Alert title Possible cause Recommended actions
A008 Density Overrange The line density is greater than
3 g/cm
3
(3000 kg/m
3
).
If other alerts are present, resolve those
alert conditions first. If the current alert
persists, continue with the recommen-
ded actions.
Check for two-phase flow. See
Section 10.23.
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
Check the drive gain and the pickoff
voltage. See Section 10.24 and
Section 10.25.
Perform Known Density Verification.
Contact Micro Motion.
A009 Transmitter Initializ-
ing/Warming Up
Transmitter is in power-up mode. Allow the meter to complete its power-
up sequence. The alert should clear au-
tomatically.
If other alerts are present, resolve those
alert conditions first. If the current alert
persists, continue with the recommen-
ded actions.
Verify that the transmitter is receiving
sufficient power.
- If it is not, correct the problem and
cycle power to the meter.
- If it is, this suggests that the trans-
mitter has an internal power issue.
Replace the transmitter.
A010 Calibration Failure Many possible causes.
This alert will not clear until you cy-
cle power to the meter.
Ensure that your calibration procedure
meets the documented requirements,
cycle power to the meter, then retry the
procedure.
If this alert appears during zeroing, veri-
fy that there is no flow through the sen-
sor, cycle power to the meter, then re-
try the procedure.
A011 Zero Calibration
Failed: Low
Many possible causes, such as too
much flow, especially reverse flow,
through the sensor during a calibra-
tion procedure, or a zero result that
is too low. This alert is accompanied
by A010.
This alert will not clear until you cy-
cle power to the meter.
Verify that there is no flow through the
sensor, cycle power to the meter, then
retry the procedure.
Troubleshooting
144 Micro Motion
®
Compact Density Meters (CDM)