Alert num-
ber Alert title Possible cause Recommended actions
A012 Zero Calibration
Failed: High
Many possible causes, such as too
much flow, especially forward flow,
through the sensor during a calibra-
tion procedure, or a zero result that
is too high. This alert is accompa-
nied by A010.
This alert will not clear until you cy-
cle power to the meter.
Verify that there is no flow through the
sensor, cycle power to the meter, then
retry the procedure.
A013 Zero Calibration
Failed: Unstable
There was too much process insta-
bility during the calibration proce-
dure.
This alert will not clear until you cy-
cle power to the meter.
Remove or reduce sources of electro-
mechanical noise (e.g., pumps, vibra-
tion, pipe stress), cycle power to the
meter, then retry the procedure.
A014 Transmitter Failure Many possible causes. Ensure that all wiring compartment
covers are installed correctly.
Ensure that all transmitter wiring meets
specifications and that all cable shields
are properly terminated.
Check the grounding of all compo-
nents. See Section 10.3.
Evaluate the environment for sources of
high electromagnetic interference
(EMI) and relocate the transmitter or
wiring as necessary.
Contact Micro Motion.
A016 Sensor Temperature
(RTD) Failure
The value computed for the resist-
ance of the line RTD is outside lim-
its.
Check your process conditions against
the values reported by the device.
Contact Micro Motion.
A017 Sensor Case Tem-
perature (RTD) Fail-
ure
The values computed for the resist-
ance of the meter and case RTDs
are outside limits.
Check your process conditions against
the values reported by the device. Tem-
perature should be between –200 °F
and +400 °F.
Contact Micro Motion.
A020 Calibration Factors
Missing
Some calibration factors have not
been entered or are incorrect.
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
A021 Transmitter/Sensor/
Software Mismatch
The configured board type does not
match the physical board.
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
Ensure that the correct board is instal-
led.
A029 Internal Electronics
Failure
This can indicate a loss of communi-
cation between the transmitter and
the display module.
Cycle power to the meter.
Replace the display module.
Contact Micro Motion.
Troubleshooting
Configuration and Use Manual 145