Option Description
Informa-
tional
Actions when fault is detected:
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
The status LED returns to green.
Status alerts and options for Status Alert Severity
Status alerts and Status Alert SeverityTable 5-1:
Alert number Alert title Default severity User can reset severity
A001 EEPROM Error Fault No
A002 RAM Error Fault No
A003 No Sensor Response Fault Yes
A004 Temperature Overrange Fault No
A005 Velocity Overrange Fault Yes
A006 Characterization Required Fault Yes
A008 Density Overrange Fault Yes
A009 Transmitter Initializing/Warm-
ing Up
Ignore Yes
A010 Calibration Failure Fault No
A011 Zero Calibration Failed: Low Fault Yes
A012 Zero Calibration Failed: High Fault Yes
A013 Zero Calibration Failed: Unsta-
ble
Fault Yes
A014 Transmitter Failure Fault No
A016 Sensor Temperature (RTD)
Failure
Fault Yes
A017 Sensor Case Temperature
(RTD) Failure
Informational Yes
A020 Calibration Factors Missing Fault Yes
A021 Transmitter/Sensor/Software
Mismatch
Fault No
A029 Internal Electronics Failure Fault No
A030 Incorrect Board Type Fault No
A033 Insufficient Pickoff Signal Fault Yes
A037 Sensor Check Failed Fault Yes
A038 Time Period Signal Out of
Range
Fault No