Analysis Subsystem 3Flex Service Manual
2-12 Aug 2013
The vacuum valve (7) should be kept closed when dosing gas in the manifold to maintain cleanliness
of the gauge. Open Valve 7 only when the system is under vacuum.
Right-click on the vacuum gauge reading to view vacuum gauge details (you must be in Service Test
mode). The combined reading is what you see displayed on the schematic.