17Stirred Ultraltration Cells
Operating Guidelines
See membrane instructions for cleaning and storage
procedures, and for buffer and solvent compat-
ibility.
Highly viscous solutions lter slowly, as do solutions
containing particulate matter, such as colloids.
Where a viscous agent (sucrose, glycerin, etc.)
is to be removed, ow can often be increased by
pre-dilution.
Prelter or centrifuge any solution containing par-
ticulate matter, such as cell debris or precipitates.
During extended dialtration, liquid level may
increase slightly. To correct, concentrate briey.
To maximize recovery of retained substances,
continue stirring for a few minutes after
depressurization. This will resuspend the polarized
layer at the membrane surface.
CAUTION: Always check that the pressure-relief
valve knob is set to the horizontal (open)
position, before removing the cell top.
Removal of the top with the pressure-
relief valve knob in the vertical (closed)
position can create a partial vacuum,
which can rupture the membrane.