23Stirred Ultraltration Cells
Troubleshooting
Symptom: Little or No Filtrate Obtained
Rotate the pressure-relief valve knob to check for
pressure. If the cell is not pressurized, check the
nitrogen source and regulator.
Make sure the glossy side of the disc membrane faces
up.
If the sample solution is highly viscous, due to microso-
lutes, either dilute or dialter to increase ow rate.
Check the membrane holder and ltrate port for
blockage.
Symptom: Filtration Rate Abnormally High
Check the membrane for lesions, scratches and rough-
ness.
Make sure that the correct membrane type is being used.
Make sure cell is pressurized prior to placing onto
magnetic stirring table.
Check the stirrer assembly to ensure that the stirring
bar is not contacting the membrane surface.
Symptom: Cell Leaks
Be sure that the lower O-ring rests entirely on the
peripheral surface of the membrane.
Check the O-rings for nicks or cuts.
Make sure that the membrane support is seated
properly, and that the base is screwed in rmly.
Make sure that the O-rings have not been not squeezed
out of their slots.
Check the ttings on the gas inlet tubing for correct
order and position. See the Gas Pressure Line
Installation section of this user guide.