User Manual
Elix® 20, 35, 70 and 100

Table of Contents

Questions and Answers

  • J
    Jeffrey BuckSep 7, 2025
    How to resolve low pump pressure on Millipore Elix 70?
    • T
      Taylor WilsonSep 7, 2025
      If your Millipore Water Filtration System is showing 'LOW PUMP PRESSURE', try increasing the pump pressure. If the issue continues, contact Millipore SAS for assistance.
  • J
    joannemontgomeryAug 22, 2025
    How to clear EXCH. ASM UV lamp in XX days message on Millipore Elix 70?
    • G
      grossmasonAug 22, 2025
      The message 'EXCH. ASM UV LAMP IN XX DAYS' on your Millipore Water Filtration System will disappear after you install a new ASM UV Lamp and reset its software timer.
  • A
    alyssa09Sep 4, 2025
    How to fix QGard TL polisher not in place error on Millipore Water Filtration Systems?
    • D
      Douglas StoneSep 4, 2025
      To resolve the 'QGARD TL POLISHER NOT IN PLACE' error on your Millipore Water Filtration System, verify that the QGard TL Polisher is installed with its “TAG” Sensor at the top.
  • J
    jeffrey65Sep 3, 2025
    How to fix Progard TL pack not in place error on Millipore Elix 70?
    • T
      Timothy MorrisSep 3, 2025
      To resolve the 'PROGARD TL PACK NOT IN PLACE' error on your Millipore Water Filtration System, make sure the Progard TL is installed with its “TAG” at the bottom.
  • H
    Henry JohnsonSep 2, 2025
    What to do if product conductivity is less than mini on Millipore Water Filtration Systems?
    • Q
      qmayoSep 2, 2025
      If the Millipore Water Filtration System displays 'PRODUCT CONDUCTIVITY < MINI', allow the Elix System to remain in PRODUCTION mode for a few minutes. This will help remove any trapped air in the Resistivity Sensor. For further assistance, contact Millipore SAS.
  • J
    Jasmine RayAug 31, 2025
    How to clear EXCH. Progard TL in XX days message on Millipore Elix 70 Water Filtration Systems?
    • L
      leslieparkerSep 1, 2025
      The 'EXCH. PROGARD TL IN XX DAYS' message on your Millipore Water Filtration System will disappear once you install a new Progard TL Pack.
  • M
    Mary PhillipsAug 30, 2025
    What to do if Millipore Elix 70 Water Filtration Systems says Progard TL pack exhausted?
    • G
      gtorresAug 30, 2025
      If your Millipore Water Filtration System displays 'PROGARD TL PACK EXHAUSTED', you should replace the Progard TL pack. Alternatively, you can accept the used-up Progard until a new one becomes available.
  • D
    David MartinezAug 29, 2025
    How to fix % rejection < setpoint on Millipore Elix 70?
    • G
      Gregory SullivanAug 29, 2025
      If the Millipore Water Filtration System is showing '% REJECTION < SETPOINT', verify that the RO % Rejection Setpoint is at 92%. Allow the Elix System to run for a few minutes, as this may increase the measured RO % Rejection.
  • R
    Rachel JonesAug 27, 2025
    What to do if the Millipore Elix 70 Water Filtration Systems shows a low feed pressure?
    • S
      smithjamesAug 28, 2025
      If the Millipore Water Filtration System displays 'LOW FEED PRESSURE', first ensure the feedwater supply is turned on. Next, check if the Progard TL Pack is clogged by observing the RO reject water in FLUSH mode. If the water pressure issue is resolved during FLUSH mode, the alarm will automatically reset.
  • A
    Anna PetersenAug 27, 2025
    What to do if distilled water conductivity is less than mini on Millipore Elix 70 Water Filtration Systems?
    • C
      candice42Aug 27, 2025
      If the Millipore Water Filtration System displays 'DIST WATER CONDUCTIVITY < MINI', keep the loop operating for a few minutes to remove any air that could be trapped in the Resistivity Sensor. If problems persist, contact Millipore SAS for assistance.

Summary

Chapter 1 Introduction

1-3 Safety Information

Details important safety statements, warnings, and symbols for operating the system.

Chapter 3 Using the Elix System

Chapter 4 Maintenance

4-1 Scheduled Maintenance Table

Provides a table outlining typical maintenance tasks, their frequency, and how to perform them.

4-2 How to Replace the Progard TL

Step-by-step instructions for removing the old and installing a new Progard TL pretreatment pack.

4-3 How to Replace the UV Lamp

Instructions for replacing the UV lamp, noting it is often recommended to be done by a service representative.

4-5 How to Sanitize the RO Cartridge(s)

Explains both automatic and manual methods for sanitizing the RO cartridges.

4-6 How to Clean the RO Cartridge(s)

Details the process for cleaning RO cartridges, including necessary kits and cleaners.

4-7 How to Replace the QGard TL1 Polisher Pack

Step-by-step instructions for replacing the QGard TL1 Polisher Pack.

Chapter 5 Troubleshooting

5-1 Maintenance, Alarm, Alarm Stop, and Custom Messages

Classifies and explains the different types of messages displayed by the Elix system.

5-2 An Alarm or Maintenance Message Is Displayed

Lists specific LCD messages, their meanings, and recommended actions for alarms and maintenance.

Millipore Elix 70 Specifications

General IconGeneral
Product Water TOC< 30 ppb
Feed Water TypePotable water
Purification TechnologyReverse Osmosis, Electrodeionization
Output Water TypeType 2 Water
Bacteria<1 CFU/mL
Bacteria Retention>99.9%
Particle Retention>99.9%
Power Supply50/60 Hz
DisplayLCD
TypeLaboratory Water Purification System