Chemical and technical specifications
42 Elix® Essential 3, 5, 10, 15 (UV) / RiOs™ Essential 5, 8, 16, 24
Specifications and requirements
Product water specifications
Ion rejection (RiOs™ Essential) > 95%
Particle rejection (RiOs™ Essential) > 99%
Resistivity (Elix® - Elix® UV)
> 5 M.cm @ 25 °C
Conductivity (Elix® - Elix® UV) < 0.2 µS/cm @ 25 °C
Total Organic Carbon (TOC) (Elix® - Elix® UV) < 30 ppb
Micro organisms for Elix® < 100 cfu/mL
Micro organisms for Elix® UV < 10 cfu/mL
Dissolved silica
< 5 ppb
System and reservoir max length tubing 6 m
Product water flow rates
Elix® Essential 3 (UV) system 3 L/h
Elix® Essential 5 (UV) system 5 L/h
Elix® Essential 10 (UV) system 10 L/h
Elix® Essential 15 (UV) system 15 L/h
RiOs™ Essential 5 system 5 L/h
RiOs™ Essential 8 system 8 L/h
RiOs™ Essential 16 system 16 L/h
RiOs™ Essential 24 system 24 L/h
Feed water quality requirements
Pressure 1 6 bar
Flow rate > 5 L/min at 2 bar
Type Potable
Temperature 5 35 °C
Conductivity 100 2000 µS/cm at 25 °C
pH 4- 10
Langelier Saturation Index (LSI) < 0.3
Free total chlorine < 3 ppm
Water system weight
Dry weight (Elix® - Elix® UV) 13.8 - 15.5 kg
Operating weight (Elix® - Elix® UV) 17.2 - 18.7 kg
Dry weight (RiOs) 12.4 - 13.6 kg
Operating weight (RiOs™) 14.4 - 15.7 kg