System
20 Elix® Essential 3, 5, 10, 15 (UV) / RiOs™ Essential 5, 8, 16, 24
Configure system parameters
Access and configure all system parameters.
SP 01
Set point for the tank level where the water system goes in Tank
Filling.
SP 02
Set point for the RO % Rejection. Below this set point, an alarm is
indicated. This measurement indicates t
he percent of ions removed
by the RO relative to the amount of ions entering it.
SP 03
Set point for the resistivity or conductivity for product water.
Below this set point, an alarm is indicated.
DATE
Menu where date and time can be adjusted.
IP AD – used with Millitrack® software
Menu where the actual IP address can be viewed.
IP DF - used with Millitrack® software
Menu where the actual IP address can be reset and switched to
the default one.
ECO
Menu where the display energy saving mode can be activated or
deactivated.
ASM
Menu where Automatic Sanitization Module can be activated.
C - - -
Restricted menu accessible with a code and used for Service.