Other functions
Elix® Essential 3, 5, 10, 15 (UV) / RiOs™ Essential 5, 8, 16, 24 35
System is not used for a long time - Lab Closed
You can use the Lab Closed function whenever the system is not be used for a long time.
In this mode, water quality is maintained by periodic automatic flushing and rinsing
Do not power off the system! The system time and date can be lost as a
result. If the system is shut off for a long period of time, contact your
service representative.
If required, empty the tank using the
front valve.
1- Insert one end of the tubing in the ball
valve. And the other end to drain.
2- Open the ball valve.
Configuring Lab Closed mode
1- Put the system in
Standby.
2- Press on right arrow
until you reach:
3- Press
to activate the
Lab Closed mode.
Exiting Lab Closed Mode
1- Close the ball valve
2-From lab closed mode:
press
for 1 second.
3-The system goes in Tank
filling.