System
16 Elix® Essential 3, 5, 10, 15 (UV) / RiOs™ Essential 5, 8, 16, 24
Display icons
Know how to interpret each icon displayed. Display with all symbols is represented below.
Item
Description
1 System’s status icons are
• Maintenance or tool:
the system is in Standby or
maintenance can be started.
• Configuration or gears:
configuration parameters such
as set points can be changed.
• Process or hour glass:
the system is busy. No action
can be done.
2 Warning maintenance or alarm (from
left to right) are displayed to indicate
that:
• CL2 sanitization is required:
perform the sanitization of the
RO cartridge.
• System sanitization is required.
• Product water resistivity is
below set point.
• Low feed water pressure is
detected.
• Water is detected (if water
sensor installed).
• Contact Millipore.
Item
Description
3 Values area is used to display
• Water quality measures.
• Software menu and
configuration values.
• Error code numbers.
4 Units are
• ‘ or minutes.
• MΩ.cm or megOhm centimeter.
• µS/cm or micro siemens by
centimetre.
• °C or temperature degree
Celsius.
• % or percentage.
5
• Consumables status icons are
used to indicate that a
maintenance action is
required.
• For more information, refer to
the Quick Guide or the
Consumables Replacement
section in this document.
6
• Tank water level is represented
by 10 bars. Each bar represents
10% of the total volume of
water in the tank.
• The outline is blinking when
the tank is empty.