Programming And Calibration
The following parameters are in the Automatic Parameter Access (APA) loop. Press
"ENTER"
once to accept
the current parameter value and proceed to the next parameter. To alter a parameter value, key in the new
value and press
"ENTER"
. Press
"ENTER"
again to advance to the next parameter.
with
[P ]
displayed
press
"ENTER"
[P01 1]
displayed, Belt Scale Application
press
"2"
press
"ENTER"
[P01 2]
displayed, Flowmeter Application
press
"ENTER"
[P02 1] displayed, Load Cell Input
key in value, 1 = Load Cell
2 = LVDT
press
"ENTER"
[P0 1]
displayed, Rate Units = MTPH
key in value, 1 = MTPH
2 = KG/H
3 = LTPH
4 = STPH
5 = LB/H
press
"ENTER"
[P1 0.000]
displayed, Design Rate = 0
key in value, Rate from Design Data Sheet or Order Acknowledgement
press
"ENTER"
[P2 1.000]
displayed, Local Totalizer Update = every 1 unit.
key in value, (refer to Parameters, P2)
press
"ENTER"
[P3 1.000] displayed, Remote Totalizer Update = every 1 unit.
key in value, (refer to Parameters, P3)
press
"ENTER"
[P7 ####]
displayed, Test Rate, 100% of Design Rate
key in value, from Design Data Sheet or refer to the associated flowmeter instruction manual.