P14 Maximum Output Adjust
Permits limited adjustment of the Compu-M maximum Analog mA Output, 20 mA. This adjustment may be
used to match the maximum input required by the attached external device. e.g. mA meter.
When this parameter is accessed the Compu-M will display a value ≅ 3300, the mA output will be 20 mA.
To increase the output press
"4"
. To decrease the output press
"8"
. The Compu-M display will increase or
decrease in relation to the change in the analog mA output.
e.g. select P14
press
"ENTER"
P14
[3325]
displayed
external meter displays 20.2 mA
press
"8"
to decrease the displayed value until 20.0 mA is displayed on the external meter.
press
"ENTER"
P15 Diagnostic Meter (V)
Activates the Compu-M input signal measurement functions. To advance to the next function press
"ENTER"
.
[P15U ###]
displayed, range = 0 to 3.98 Volts DC
The value displayed is proportional to the input level supplied by the load cell(s) or the LVDT.
[F #####]
displayed, range = 0 to 131,072 counts.
High resolution reference value display, proportional to load cell or LVDT input signal levels.
*
[P15A ####]
displayed, Load Cell A input in mV (P02=1).
*
[P15b ####]
displayed, Load Cell B input in mV (P02=1).
select P15
[P15 ]
displayed.
press
"ENTER"
[P15U ###]
displayed.
press
"ENTER"
[F #####]
displayed.
press
"ENTER"
*
[P15A ####]
displayed.
press
"ENTER"
*
[P15b ####]
displayed.
* Load Cell applications only, (P02=1)
PL-516 64