P67 Direct Zero Entry
Allows direct access to the zero reference count.
Direct entry is intended for use when replacing software or hardware and
it is not convenient to perform an initial zero at that time.
enter : 0 = disabled
1 = enabled
press "ENTER"
[Pc #####] displayed, previous count (display ’0’ if no previous count)
press "12345" e.g. count value 12345
press "ENTER"
[oC 12345] e.g. new count value
P68 Direct Span Entry
Allows direct access to the span reference count.
Direct entry is intended for use when replacing software or hardware and it is not convenient
to perform a span at that time to establish a proper initial span.
enter: 0 = disabled
1 = enabled
press "ENTER"
[Pc #####] displayed, previous count (display ’0’ if no previous count)
press "67890" e.g. count value 67890
press "ENTER"
[SC 67890] e.g. new count value
P77 Initial Zero
Activates a Zero calibration which the Compu-M will utilize as a reference to which subsequent routine Zero
calibrations will be compared.
This parameter is generally utilized in response to a Display Message
[E3]
, when an accountable but
mechanically non recoverable zero deviation, greater than the zero limit (P25), has occurred.
Refer to the appropriate recalibration section of this manual.
enter:
"1"
, Initial Zero (F = 0)
PL-516 69