7ML19981CJ01 MiniRanger Plus Page 7
Specifications
Specifications
MiniRanger Plus
Power
• 9-250 Vac, 40-70 Hz, 38 VA, 12W or 9-250 Vdc, 11W
• leakage current to ground: 3 mA at 120 Vac, 3.5 mA at 240 Vac
Fuse
• F1 2AG, FastBlow, 1.5 amp, 250 V
• F3 2 AG, SlowBlow, 1 amp, 250 V
• F4 Belling Lee, L754, 4000A HRC, ceramic type 100 mA, 250V
Environmental
• wall / rack / panel specific
Range
• solids applications: 7.5 m (25 ft.) (transducer dependent)
• liquid applications: 15 m (50 ft.) (transducer dependent)
Accuracy
• 0.25% of range or 6 mm (0.24"), whichever is greater
Resolution
• 2 mm
Memory
• program stored in non-volatile FLASH memory, for software upgrades
via optional Dolphin interface
• parameters stored in non-volatile EEPROM
Display
• wall / rack / panel specific
Outputs
Transducer Drive
• 43.5 kHz, 250 V peak pulses of 1 ms max. duration at max. repetition
rate of 300 ms
mA
• optically isolated 0/4 - 20 mA
• max. loading: 750 Ω
• resolution: 0.1% of 20 mA
Relays
• 2 alarm/pump control relays
• 1 form 'C' SPDT contact per relay, rated 5 A at 250 Vac non-inductive
Communication
• wall / rack / panel specific
Approvals
• CE
1
, FM, CSA
NRTL/C
1
EMC performance available upon request.