6 - 2
• Always input the correct patient information before you start 12-Lead ECG analysis because the
patient information, especially age, gender, and race, greatly affect the interpretation of the
acquired ECG.
• If no patient information is inputted, the system will interpret the acquired ECG on the basis of a
default of a 50 years old Caucasian male, which may result in misdiagnosis.
• The Lead Select hard key is disabled in the 12-lead ECG screen.
• During 12-lead ECG acquisition or analysis, you cannot select a parameter area to enter the
parameter setup menu. The Main Menu button and the Event button on the front panel are also
disabled.
• The filter mode is set to [ST] automatically for 12-lead ECG acquisition and analysis.
• If the defibrillator is on a moving vehicle, stop the vehicle when acquiring 12-lead ECG.