10 - 1
10 Noninvasive Pacing
10.1 Pacing Introduction
• In the Pacer mode, arrhythmia analysis is supported and available arrhythmia alarms are asystole,
ventricular vibrillation and ventricular tachycardia.
10.2 Pacing Safety Information
• Heart rate displays and alarms function during pacing, but they can be unreliable. Observe the
patient closely while pacing. Do not rely on the indicated heart rate or heart rate alarms as a
measure of the patient’s perfusion status.
• To avoid explosion hazard when pacing a patient who is receiving oxygen delivery, properly route
the oxygen delivery tube. Do not keep it close to the electrode pads.
• Monitoring ECG alone is sometimes not enough to verify that the patient’s heart is providing cardiac
output.a patient’s response to pacing shall be verified by signs of improved cardiac output, such as:
a palpable pulse rate the same as the rate which pace pulses are being delivered, a rise in blood
pressure, and/or improved skin color.
• Use of Pacer mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of
pacing therapy.
• For treatment of patients with implanted devices such as permanent pacemakers or cardioverter-
defibrillators, consult a physician and the instructions for use provided by the device’s manufacturer
• Prolonged noninvasive pacing may cause patient skin irritation and burns. Periodically inspect the
underlying skin and change ECG electrodes and electrode pads.
• If pacing is interrupted for any reason, the [Start Pacing] soft key must be pressed to resume pacing.
• In the Pacer mode, you cannot change the patient’s internal paced status from the [ECG Setup]
menu.
• In the case that electrode pads poorly contact the patient, the alarm “Pacer Stopped Abnormally”
and “Pads Off” may be presented.
• Electrode pads are not an available choice for the source of ECG waveform in the Pacer mode.