21 Network Connection
21.1 Network Safety Information
CAUTION
Wireless network design, deployment, debugging, and maintenance should be executed by Mindray
service personnel or authorized technicians.
Always deploy the wireless network according to local wireless regulations.
To ensure network security and stability, data communication should be performed within a closed
network or within a virtually isolated network provided by a hospital for all network functions. The
hospital is responsible for ensuring the security of the virtually isolated network.
Keep network authentication information, for example password, from being accessed by
unauthorized users.
Do not connect non-medical devices to the hospital network.
If wireless network signal is poor, there may be a risk of CMS data loss.
RF interference may result in wireless network disconnection.
Disconnecting from the network may result in CMS data loss and function failure. Check the patient
in case of network disconnection and reconnect the network problem as soon as possible.
Ensure that the equipment network settings are correct. Changing the network settings may result
in network disconnection. Contact your service personnel if you have any problems on the network
settings.