A - 11
A.5.5 PR Specifications
PR from Mindray SpO
2
Module
PR from Masimo SpO
2
Module
PR from Nellcor SpO
2
Module
PR from IBP Module
A.5.6 Temp Specifications
A.5.7 NIBP Specifications
Measurement range 20 to 300 bpm
Resolution 1 bpm
Accuracy ±3 bpm
Response time <20 s (PR value sudden changes from 25 to 240 bpm)
Measurement range 25 to 240 bpm
Accuracy ±3 bpm (measured without motion)
±5 bpm (measured with motion)
Response time ≤20 s (PR value sudden changes from 25 to 220 bpm)
Measurement range 20 to 300 bpm
Resolution 1 bpm
Accuracy ±3 bpm (20 to 250 bpm)
Not specified (251 to 300 bpm)
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Standard Meet the standard of ISO 80601-2-56
Operating mode Direct mode
Parameters Max. 2 channels. T1, T2, TD
Measurement range 0 to 50 °C (32 to 122 °F)
Accuracy ±0.1 °C or ±0.2 °F (without considering probe error)
Resolution 0.1 °C
Minimum time for accurate measurement Body surface: <100 s
Body cavity: <80 s
Standards Meet standard of IEC 80601-2-30
Technique Oscillometr y
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time 5 min
Static pressure measurement range 0mmHg to 300mmHg