14.3.2 Starting Auto NIBP Measurements
WARNING
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and
neuropathy in the limb with the cuff. Inspect the application site regularly to ensure skin quality and
inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If any abnormity
occurs, move the cuff to another site or stop the blood pressure measurements immediately.
14.4 Setting Initial Cuff Inflation Pressure
NOTE
For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the
measurement time.
Setting initial cuff inflation pressure is disabled during NIBP measurement.
The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or
patient category has been changed.