25 - 1
25 Maintenance
25.1 Maintenance Safety Information
• Failure for the responsible individual, hospital or institution employing this equipment to
implement a satisfactory maintenance schedule may cause undue equipment failure and possible
health hazards.
• The safety checks or maintenance involving any disassembly of the equipment should be performed
by professional servicing personnel. Otherwise, undue equipment failure and possible health
hazards could result.
• If you find a problem with any of the equipment, contact your service personnel or the
manufacturer.
• No modification of this equipment is allowed.
• The service personnel must be properly qualified and thoroughly familiar with the operation of the
equipment.
• Do not touch any connected electrode pads or external paddles with hands during user test and
auto test. Otherwise, electric shock could result.
• Do not open the equipment housings. All servicing and future upgrades must be carried out by the
service personnel.
25.2 Routine Maintenance
25.2.1 Auto Test
• If turned off, the auto test is performed only when AC mains is connected.
• The auto test simulates the discharge test through impedances in the paddle tray. The auto test
passes only when external paddles properly contact the metal parts of the paddle tray.
• Thoroughly clean the external paddles and properly place them in the paddle tray after each use.
Otherwise, the auto test may fail or damaged external paddles may result.
• The auto test reduces the battery power. If the equipment is not connected to the AC mains
immediately, low battery may result.
• Before the auto test, check that the equipment is connected to the AC mains with a battery installed,
and external paddles are properly placed in the paddle tray or the equipment is connected with the
pads cable and 50 Ω test load. If the pads cable is not connected with the 50 Ω test load, the message
“Test load not connected with cable” appears when the auto test passes. This means that the
equipment only passes the internal discharge test, but not pass the external discharge test
connected with the test load.
25.2.2 User Test
• Do not perform the user test when a patient is connected to the equipment.