A.3 CPR Compression Specifications
CPR Compressions from Defibrillation Electrodes
CPR Compressions from CPR Sensor
A.4 Pacer Specifications
NOTE
When pacing rate is changed from 30ppm to 210ppm, the response time to pacing (HR rising from
30bpm to 210bpm) is less than 20s.
Compression rate Measurement range: 60 to 200 cpm (compressions per minute)
Accuracy: ±3 cpm (compression per minute)
Compression depth Measurement range: 0.0 to 8.0 cm
Effective range: 1.5 to 8.0 cm
Accuracy (for effective range): ±0.5 cm or ±10%, whichever is greater
Resolution: 0.1 cm
Refreshing rate: ≥0.5Hz
Compression rate Measurement range: 40 to 160 cpm (compressions per minute)
Effective range: 40 to 160 cpm (compressions per minute)
Accuracy: ±2 cpm (compression per minute)
Resolution: 1 cpm
Refreshing rate: ≥0.5Hz
Interruption time Measurement range: 0 to 300 s
Effective range: 0 to 300 s
Resolution: 1 s
Refreshing rate: ≥0.5Hz
Standards Meet standards of IEC 60601-2-4
Pacer mode Demand, fixed
Output waveform Monophasic square wave pulse
pulse width 20 ms or 40 ms
Accuracy: ±5%
Pacing rate 30ppm to 210ppm
Accuracy: ±1.5%
Resolution: 5 ppm
Pacing output 0mA to 200mA,
Accuracy: ±5% or ±5mA, whichever is greater
Resolution: 1mA, 2mA or 5mA
Refractory period 200 to 300 ms (depending on pacing rate)
4:1 pacing Pacing pulse frequency reduced by factor of 4 when this function is
activated.
Output protection The equipment has no sign of damage after defibrillation-proof test.