A - 7
Measurement range Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Accuracy ±1% or ±1bpm, which ever is greater
Resolution 1 bpm
Sensitivity 200 V (lead II)
Heart rate averaging In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC
60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most
recent RR intervals are averaged to compute the HR. Otherwise, heart rate is
computed by subtracting the maximum and minimum ones from the most
recent 12 RR intervals and then averaging them.
The HR value displayed on the screen is updated every second.
Response time to heart rate change Meets the requirements of IEC 60601-2-27: Clause 201.7.9.2.9.101 b) 5).
From 80 to 120 bpm:
From 80 to 40 bpm:
less than 11 s
less than 11 s
Time to alarm for tachycardia Meets the requirements in Clause 201.7.9.2.9.101 b) 6) of IEC 60601-2-27.
Waveform
4ah - range:
4a - range:
4ad - range:
4bh - range:
4b - range:
4bd - range:
<11 s
<11 s
<11 s
<11 s
<11 s
<11 s
Arrhythmia Analysis Classifications Asystole, V-Fib/V-Tach, V-Tach, Vent. Brady, Extreme Tachy, Extreme Brady,
PVCs/min, PVC, Couplet, VT>2,Bigeminy, Trigeminy, R ON T, Brady, Tachy,
Missed Beat, PNP, PNC, Vent Rhythm, Multif. PVCs, Nonsus. Vtac, Pause, Irr.
Rhythm, A-Fib
Tall T-wave rejection capability When the test is performed based on Clause 201.12.1.101.17 of IEC 60601-2-
27, the heart rate meter will reject all 100 ms QRS complexes with less than
1.2 mV of amplitude, and T waves with T-wave interval of 180 ms and those
with Q-T interval of 350 ms.
Lead-off detection current Measuring electrode:
Drive electrode:
≤0.1 A
≤1 A
Baseline recovery time <2.5 s (after defibrillation)
Response to irregular rhythm In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC
60601-2-27, the heart rate after 20 seconds of stabilization is displayed as
follows:
Ventricular bigeminy (3a): 80±1 bpm
Slow alternating ventricular bigeminy (3b): 60±1 bpm
Rapid alternating ventricular bigeminy (3c): 120±1 bpm
Bidirectional systoles (3d): 90±2 bpm
12-lead ECG Interpretation
Sampling rate 500 samples/second/channel
Ampitude quantisation 1 V/LSB